Active Versus Passive Cooling During Neonatal Transport.


BACKGROUND AND OBJECTIVE: Therapeutic hypothermia is now the standard of care for hypoxic-ischemic encephalopathy. Treatment should be started early, and it is often necessary to transfer the infant to a regional NICU for ongoing care. There are no large studies reporting outcomes from infants cooled passively compared with active (servo-controlled) cooling during transfer. Our goal was to review data from a regional transport service, comparing both methods of cooling.

METHODS: This was a retrospective observational study of 143 infants referred to a regional NICU for ongoing therapeutic hypothermia. Of the 134 infants transferred, the first 64 were cooled passively, and 70 were subsequently cooled after purchase of a servo-controlled mattress. Key outcome measures were time to arrival at the regional unit, temperature at referral and arrival at the regional unit, and temperature stability during transfer.

RESULTS: The age cooling was started was significantly shorter in the actively cooled group (46 [0–352] minutes vs 120 [0–502] minutes; P <.01). The median (range) stabilization time (153 [60–385] minutes vs 133 [45–505] minutes; P = .04) and age at arrival at the regional unit (504 [191–924] minutes vs 452 [225–1265]) minutes; P = .01) were significantly shorter in the actively cooled group. Only 39% of infants passively cooled were within the target temperature range at arrival to the regional unit compared with 100% actively cooled.

CONCLUSIONS: Servo-controlled active cooling has been shown to improve temperature stability and is associated with a reduction in transfer time.

Source: http://pediatrics.aappublications.org